FortisCare EM66
|
FortisCare EM66
Meter Specifications |
Blood specimen |
Capillary whole blood/ Venous whole blood |
Memory capacity |
480 |
Blood sample |
0.6 μL Fresh blood from fingertip, palm, or forearm |
Standards |
IVD 98/79/EC
ISO 13485 : 2003
ISO 14971 : 2009
ISO 15197 : 2013
EN 980 : 2008
EN 61326-1 : 2006
EN 61326-2-6 : 2005
EN 61010-2-101 : 2002
|
Test time |
5 seconds |
AST |
Palm and Forearm |
HCT range |
25~55% |
Measuring range |
20~600 mg/dL |
Accuracy |
95% within + 15 mg/dL when glucose <100mg/dL
95% within + 15% when glucose >100mg/dL |
Precision |
CV (correlation variation) is less than 5 % |
Storage & Transportation Temp. |
35.6°F -86°F (2°C -30°C) |
Temperature and Humidity |
Operating : 10℃~40℃, below 85% R.H.
Meter storage : -10℃~65℃, below 90% R.H.
Strip storage : 2℃~30℃, below 90% R.H.
Transportation : 2℃~30℃, below 90% R.H.
|
How calibrated |
No code |
Meter dimension |
83.7 x 51.2 x 22.9mm |
Power Supply |
Alkaline Battery 1.5v x2 (AAA) |
Battery life |
Over 1000 tests |
Data port |
USB |
Auto shut-off |
Power off functions after 90 seconds no key presse |
Time Setting |
12 Hours or 24 Hours (Optional) |
Principles |
Electrochemical biosensor technology |
Display |
Liquid crystal display module
With inverse backlight
Dimension : 56.8 x 37 mm
|
Error Codes |
Err : humidified / Used strips
LP : Low power, please replace new battery
001 with err icon : Not enough blood
005 with err icon : system fault
HI : test result is higher than 630 mg/dL
LO: test result is lower than 20 mg/dL
Ht : Temperature too high to have a test
Lt : Temperature too low to have a test
|
|
|
|